A Fully Magnetically Levitated Left Ventricular Assist Device — Final Report
Material type: ArticleBoston: Massachusetts Medical Society, 2019Description: 1618 - 1627Subject(s): DDC classification:- 21
Item type | Current library | Collection | Call number | Vol info | Status | Date due | Barcode | |
---|---|---|---|---|---|---|---|---|
Hemeroteca | Biblioteca Gustavo Vásquez Betancourt. Sede Medellín Torre 1 | Colección Ciencias de la Salud | Vol 380, No 17 (25/04/2019) -- Vol 380, No 17 (25/04/2019) | Available | 20312 |
BACKGROUND
In two interim analyses of this trial, patients with advanced heart failure who were treated with a fully magnetically levitated centrifugal-flow left ventricular assist device were less likely to have pump thrombosis or nondisabling stroke than were patients treated with a mechanical-bearing axial-flow left ventricular assist device.
METHODS
We randomly assigned patients with advanced heart failure to receive either the centrifugal-flow pump or the axial-flow pump irrespective of the intended goal of use (bridge to transplantation or destination therapy). The composite primary end point was survival at 2 years free of disabling stroke or reoperation to replace or remove a malfunctioning device. The principal secondary end point was pump replacement at 2 years.
RESULTS
This final analysis included 1028 enrolled patients: 516 in the centrifugal-flow pump group and 512 in the axial-flow pump group. In the analysis of the primary end point, 397 patients (76.9%) in the centrifugal-flow pump group, as compared with 332 (64.8%) in the axial-flow pump group, remained alive and free of disabling stroke or reoperation to replace or remove a malfunctioning device at 2 years (relative risk, 0.84; 95% confidence interval [CI], 0.78 to 0.91; P<0.001 for superiority). Pump replacement was less common in the centrifugal-flow pump group than in the axial-flow pump group (12 patients [2.3%] vs. 57 patients [11.3%]; relative risk, 0.21; 95% CI, 0.11 to 0.38; P<0.001). The numbers of events per patient-year for stroke of any severity, major bleeding, and gastrointestinal hemorrhage were lower in the centrifugal-flow pump group than in the axial-flow pump group.
CONCLUSIONS
Among patients with advanced heart failure, a fully magnetically levitated centrifugal-flow left ventricular assist device was associated with less frequent need for pump replacement than an axial-flow device and was superior with respect to survival free of disabling stroke or reoperation to replace or remove a malfunctioning device. (Funded by Abbott; MOMENTUM 3 ClinicalTrials.gov number, NCT02224755.)